UAT 103 Fundamentals of Electrical and Electronics

Worksheet 5 Using instruments and electrical safety (practical assessment)

Lab 6 hAssesses CLO3

Objective

Show correct and safe use of the multimeter and oscilloscope, find a circuit fault and charge a lithium battery by procedure

Equipment and materials

  • Multimeter
  • Oscilloscope
  • Lithium battery with balance lead
  • LED circuit and small-motor circuit
  • Servo tester
  • Sample harness with a fault set by the lecturer
  • Balance charger, fire-resistant bag or box and charging checklist

Safety

  • Move the lead back to the V jack after every current measurement
  • Beware the scope ground; connect the probe ground only to circuit ground
  • Charge in a fire-resistant bag or box, always attended, and never charge a swollen battery

Steps

  1. Measure each cell through the balance lead, record it and state the battery's condition
  2. Measure current in the LED and small-motor circuits, then return the lead to the V jack
  3. Measure the pulse width and frequency of the servo tester's signal
  4. Check continuity in the harness and find the fault set by the lecturer
  5. Demonstrate battery charging using the checklist

Record of results

Table 1 Station 1 cell voltages through the balance lead
CellVoltage (V)Meter uncertainty (±V)Condition
Table 2 Stations 2 and 3 current and signal
MeasurementConnection or instrument settingReadingUnitLead back in V jack
Station 4 Fault location and the order of checks used

Questions

  1. If two cells in Table 1 differ by 0.03 V and the meter is specified ±(0.5% + 2 digits), can you conclude the pack is unbalanced? Show the calculation.

  2. If you forgot to move the lead from the A jack and then measured across the battery, what would happen? Explain using the meter's resistance in current mode.

  3. Explain the order of checks you used to find the harness fault, and why continuity and resistance must be measured with the power off.

  4. For a 4S 5000 mAh pack, what charge current should be set at 1C, at what voltage per cell and for the pack does the constant-voltage stage hold, and what must be checked before charging starts?

Check before you hand in

What to hand in

  • Record sheet for all five stations
  • Fault location and how it was found
  • Charging checklist signed by the examiner

CLO3 Use electrical instruments correctly and safely. (CLOs are a web-team draft; the course owner sets the final ones.)

Instructors: answers and teaching notes for this worksheet are in the teaching pack